Abstract

A blood mimicking fluid (BMF) has been developed for the acoustic and thermal characterizations of high intensity focused ultrasound (HIFU) ablation devices. The BMF is based on a degassed and de-ionized water solution dispersed with low density polyethylene microspheres, nylon particles, gellan gum, and glycerol. A broad range of physical parameters, including attenuation coefficient, speed of sound, viscosity, thermal conductivity, and diffusivity, were characterized as a function of temperature (20-70 degrees C). The nonlinear parameter B/A and backscatter coefficient were also measured at room temperature. Importantly, the attenuation coefficient is linearly proportional to the frequency (2-8 MHz) with a slope of about 0.2 dB cm(-1) MHz(-1) in the 20-70 degrees C range as in the case of human blood. Furthermore, sound speed and bloodlike backscattering indicate the usefulness of the BMF for ultrasound flow imaging and ultrasound-guided HIFU applications. Most of the other temperature-dependent physical parameters are also close to the reported values in human blood. These properties make it a unique HIFU research tool for developing standardized exposimetry techniques, validating numerical models, and determining the safety and efficacy of HIFU ablation devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.