Abstract

A high-sensitivity bioaerosol monitor based on a single 365 nm LED is developed and a calibration approach is discussed for the first time. The fluorescence detection system, which is the core part of the monitor, contains an optical detection module and a fluorescence signal processor configured with a phase sensitive detector (PSD). B800 fluorescent microspheres and staphylococcus are used for performance evaluation of the monitor. B800 microspheres are appropriate as calibration material. The experimental results demonstrate the PSD plays a significant role in improving sensitivity and signal-to-noise ratio (SNR) of detection. Our monitor can detect staphylococcus concentration above 1800 cfu/L of air stably.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.