Abstract

A novel extraction procedure coupled with gas chromatography–mass spectrometric detection for quantification of organochlorine pesticides (OCPs) in water is described. Amphiphilic polyhydroxylated polyparaphenylene (PH-PPP) was synthesized and coated on the surfaces of a porous polypropylene hollow fiber membrane (HFM). Due to the high porosity of the HFM, maximum active surface area to achieve high extraction efficiency is expected. The polymer-coated HFM was used for the extraction of 15 OCPs from water. The extraction efficiency was compared with emerging and established methods such as liquid-phase microextraction (LPME), solid-phase microextraction (SPME) and stir bar sorptive extraction (SBSE) techniques. We term the current procedure as polymer-coated hollow fiber microextraction (PC-HFME). PC-HFME showed good selectivity and sensitivity. Detection limits for OCPs were in the range of 0.001–0.008 μg l −1. The sensitivity and selectivity of the coated HFM could be adjusted by changing the characteristics of the coated PH-PPP film.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.