Abstract

A high performance liquid chromatography (HPLC) method based on cation exchange separation has been developed for the measurement of dopamine (DA), 5-hydroxytryptamine (5-HT) and norepinephrine (NE) in microdialysates. The separation conditions have been optimised for using electrochemical detection. All three bioamines were resolved in less than 22 min using isocratic conditions. The optimum oxidation potential for the three bioamines was found to be +0.4 V vs. in situ Ag/AgCl reference electrode. Linear regression analysis of HPLC-peak area as a function of concentrations in the range 1–50 ng ml −1 gave coefficients of correlation between 0.998 and 0.999. The limit of detection for DA, 5-HT and NE was found to be between 50 and 100 pg ml −1 with a signal to noise ratio of 3:1. The method has been applied to the simultaneous measurement of the three monoamines in microdialysates from the medial prefrontal cortex under basal conditions and following the administration of the antipsychotic drug clozapine (10 mg kg −1 s.c.).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.