Abstract
Many finite element (FE) models have been designed based on geometric information from computed tomography (CT) data, and validated via comparison with experimental results for human cadaver ossicular bones. Here, we describe a novel method for developing and analyzing the biomimetic ceramic ossicles (BCO) in combination with 3D printing technology, and we establish an FE model of the BCO for analyzing vibration performance. Novel biomimetic ceramic ossicles (BCO) made of hydroxyapatite (HA) were fabricated using 3D printing technology, and their vibration properties were measured. We created a 3D model of the BCO using computer-aided design, which corresponds to the ossicular structure and geometry, and created an FE model of the human ossicles via a comparison of experimental and simulated vibrations to investigate the characteristics of the ossicular chain. The FE model was established based on the displacements of the malleus, incus, and stapes, which was analyzed using an externally applied vibrational force.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Precision Engineering and Manufacturing
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.