Abstract
A simple, accurate, rapid and precise isocratic reversed phase high-performance liquid chromatographic method has been developed and validated for determination of Doravirine in tablets. The chromatographic separation was carried out on Dionex C18 (250 x 4.6mm, 5µ) with a mixture of methanol: 0.05M potassium dihydrogen phosphate (40:60%v/v) as a mobile phase at a flow rate of 1.5 mL/min. UV detection was performed at 306 nm. The retention time was 5.24 min for Doravirine. Calibration plot was linear (r2=0.999) over the concentration range of 200-600 µg/mL. The method was validated for accuracy, precision, specificity, linearity, robustness, LOD and LOQ. The proposed method was successfully used for quantitative analysis of tablets. No interference from any component of pharmaceutical dosage form was observed. Validation studies revealed that method is specific, rapid, reliable, and reproducible. The high recovery and low relative standard deviation confirm the suitability of the method for routine determination of Doravirine in bulk and tablet dosage form.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Asian Journal of Research in Medical and Pharmaceutical Sciences
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.