Abstract

Diagnostic reference level (DRL) is an appropriate instrument toward promoting radiation doses optimisation in medical imaging. The goal of this research is developing DRL to optimise computed tomography (CT) doses in patient examination. Parameters were collected in CT facilities for common procedures such as head, chest, pelvic and cervical spine (c-spine) imaging. The dose descriptors considered were volume computed tomography dose index (CTDIv) and dose length product (DLP). The DRLs were proposed at 75th percentile CTDIv for head (without and with contrast materials), chest (without and with contrast materials), pelvic and c-spine only without contrast materials; their values were 52, 52, 17, 14, 14 and 38mGy, respectively. Whereas, DLP values for the aforementioned protocols were 1237, 1459, 625, 565, 605 and 1106mGy.cm, respectively. This study fruitfully developed the DRLs for head, chest, pelvic and c-spine and can be accepted for clinical purposes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.