Abstract

This paper proposes an all-in-one microextraction-based protocol capable of determining and quantifying fentanyl, methadone, and zolpidem in plasma, urine, and saliva at concentrations below those required by international regulatory organizations. A homemade thin-film microextraction device featuring an octyl-cyanopropyl stationary phase was coupled with LC-MS/MS. The proposed method was developed and validated according to FDA criteria, providing extraction efficiency values ranging from 26.7% to 76.2% with no significant matrix effects (2.6% to 15.5% signal suppression). The developed protocol provided low limits of quantification (mostly equal to 1 ng mL-1) and good reproducibility (intra- and inter-day RSDs of less than 9.6% and 12.0%, respectively) and accuracy (89% to 104% of the test concentration). An assessment of the protocol's environmental impact indicated that attention must be devoted to eliminating the use of toxic reagents and developing its capability for in situ sampling and in-field analysis using portable instruments. The proposed TFME-based protocol provides clinical laboratories with a versatile, one-step tool that enables the simultaneous monitoring of fentanyl, methadone, and zolpidem using the most popular biological matrices.