Abstract

AbstractCoal's calorific value and carbon content are crucial in calculating carbon emissions. Accurately detecting these two indicators is of great significance for carbon accounting. In this study, we developed a compact coal quality rapid detection integrated machine based on laser‐induced breakdown spectroscopy (LIBS), which can directly measure coal particle flow. A partial least squares model, based on data set selection according to cluster analysis results, was applied to establish the relationship between coal quality and plasma spectra. The R2 of the calorific value is .93, the root mean square error of prediction (RMSEP) is 0.41 MJ/kg, and the mean absolute error (MAE) is 0.33 MJ/kg. The R2 of carbon content is .94, the RMSEP is 0.97%, and the MAE is 0.91%. These results indicated that the developed compact coal quality rapid detection integrated machine could conduct precise coal quality analysis in real time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.