Abstract

In contrast to the articulating and taper surfaces of failed total hip replacements, volumetric wear analysis of trunnions is not routinely performed. Metal wear particles from the trunnion may contribute not only to the failures of metal-on-metal total hip replacements but also to all hip replacements utilising metal trunnions. A validation study was performed with the material removed in stages from the trunnions of an Exeter V40 stem, a Corail stem and an Accolade stem to simulate different magnitudes of wear. The material loss from the trunnions was measured both volumetrically with a coordinate measuring machine and gravimetrically with a high-precision balance. A cohort of 28 ex vivo trunnions was also measured using the coordinate measuring machine. The maximum error between the two methods was found to be 0.13 mm(3). This result was comparable with the coordinate measuring machine method for the taper surface (0.2 mm(3)). The ex vivo trunnions had a median wear volume of 0.14mm(3) (range: 0.04-0.28 mm(3)). This is the first study to determine the accuracy of volumetric wear measurements of trunnions by comparing against gravimetric measurements. Volumetric wear analysis of trunnions may provide additional insights into failures of modular total hip prostheses and will be performed routinely at our centre.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.