Abstract

Tritiated water will be present in many demonstrator class fusion devices in a wide array of locations and concentrations. Accurate and fast measurement will be a key requirement to ensure suitable process monitoring, safety assurance and tritium tracking is in place. This work quantifies the deuterium detection limits achievable using a benchtop transmission Fourier transform infrared (FTIR) instrument and an industrialised attenuated total reflectance (ATR)-FTIR instrument to understand the capabilities of these techniques for real-time monitoring of aqueous tritium. Deuterium limits of detection of 4.62 × 10-5 mol mL-1 and 1.07 × 10-3 mol mL-1 were demonstrated with measurement times of 10s for the transmission FTIR and ATR-FTIR, respectively. These instruments are considered viable for the measurement of high concentration tritiated water and have many potential benefits associated with their deployment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.