Abstract

The mercury (5 - 100 ng of Hg2+ and/or CH3Hg+) in drinking water (25 ml) containing 2,3-dimercaptopropane-1-sulfonate (DMPS, 15 µg) and sodium acetate buffer (1 mmol, pH=6.0) was concentrated on two Sep-Pak C18 cartridges in series. The mercury-DMPS complex was then eluted with methanol and adjusted to 2.00 ml. A portion (50 µl) was introduced into a graphite cuvette and was atomized according to a temperature program. The method detection limit (3 σ) was 0.053 μg/l; the calibration graphs were linear up to 4.00 μg/l. Accuracies of 97.4 - 99.3% were obtained when testing a synthetic natural water standard (Standard Reference Material 1641c). The proposed method could be applied to the determination of methylmercury in air with a detection limit of 1.3 ng.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.