Abstract

An electrothermal atomic absorption spectrometric procedure for zinc determination in animal tissues is first optimized and then used to measure the metal content of different tissues from normal and dystrophic mice. The procedure involves minimal manipulation of the sample to overcome the severe contamination problems normally associated with the measurement of low zinc levels. The samples (recommended amount 10 mg) are slurried in 2 ml of a 10-mM tetramethylammonium hydroxide solution containing 0.1 g/100 ml silicone antifoam. After mild heating at 60 °C for 10 min and homogenization for 1 min, the suspensions are submitted to a 10-fold dilution and then injected into the electrothermal atomizer. Calibration is carried out using aqueous standard solutions of zinc prepared in the same suspension media. The reliability of the whole procedure is verified using three certified reference materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.