Abstract

An analytical method for the determination of zilpaterol in livestock products was developed. The sample was stirred with n-hexane and n-hexane saturated acetonitrile, and zilpaterol in the sample was extracted with acetonitrile. The extract was cleaned up on a ODS cartridge column (1 g) and SCX cartridge column (500 mg). The LC separation was carried out using an Inertsil ODS-4 column and linear gradient elution with 0.1%formic acid and acetonitrile containing 0.1% formic acid as mobile phase. Detection of MS was carried out positive ion electrospray ionization mode. Average recoveries (n=5) of zilpaterol from 6 kinds of livestock products fortified at the MRLs (0.01 mg/kg) were 87.0-99.4%, and the relative standard deviations were 2.4-6.3%. The limits of quantitation were 0.01 mg/kg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call