Abstract
The "Ferris Wheel" RE Exposure System was designed by Motorola Inc. to study the long-term biological effects due to RE exposure [3]. The main goal of this research project was to characterize the "Ferris Wheel" to know how efficient and symmetrical was the exposure system in distributing the RE power among its loads. The characterization of the system was done in terms of power efficiency, SAR, Whole-Body SAR and Localized SAR. Exposure to RF sources is quantified in terms of SAR which defines the rate of electromagnetic deposition per unit mass. Determination of Whole-Body averaged SAR requires to have the actual amount of energy absorbed where as the temperature increase in the tissue material yields the Localized SAR. Calorimetric Techniques were used to characterize the "Ferris Wheel" exposure system in terms of Whole-Body SAR and efficiency. Microwave Studio and XFDTD simulation programs based on Finite Difference Time Domain method were also used to determine the Whole-Body SAR and B-field distribution in the "Ferris Wheel." The B-field distribution inside the FW was measured by using B-field probes and result was compared to that of simulated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.