Abstract

A simple, rapid and reliable method was developed for the determination of copper, nickel, iron and lead in fresh and used lubricating oil samples by flame atomic absorption spectrophotometry (FAAS). In the present study, a mixture of organic solvents containing propionic acid and iso-butylmethyl ketone (1: 1) was used to extract the metals from lubricating oil samples followed by FAAS analysis. Aqueous standard solutions can be easily employed with the proposed mixed solvent system instead of organometallic standards. The analytical results obtained by employing the proposed solvent extraction system were found to be in good agreement with the results for aqueous media obtained after the destruction of oil samples matrix. Percentage recovery studies showed 88–98% for Cu, 92–95% for Fe, 96–106% for Ni and 84–100% for Pb with relative standard deviation of 2–6%. The developed method was effectively applied to routine determination of Cu, Ni, Fe, and Pb in lubricating oil samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.