Abstract

A very simple high-performance liquid chromatographic method for determination of Vitamin A in pharmaceutical preparations without the need for saponification was developed. A reversed-phase (Nova-Pack C18, 4 μm) column was used with a mobile phase of acetonitrile-tetrahydrofuran-water (55∶37∶8) and a flowrate of 1.5 ml/min. Sample treatment only consisted of the extraction of retinol acetate content from capsules or tablets with methanol. Total extraction was achieved by shaking vigorously with the aid of magnetic stirring for three hours at room temperature. No change of solvent is necessary to introduce the sample in the chromatographic system. This method is suitable for routine quantification of Vitamin A.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.