Abstract

A high-performance liquid chromatographic (HPLC) method with fluorescence detection for the quantification of vancomycin in human plasma was developed and validated. The method includes an extraction of vancomycin by deproteinization with acetonitrile. The analyses were carried out at 258 nm as the emission wavelength while exciting at 225 nm on a reversed-phase column (30 cm x 4 mm i.d. x 10 microm Waters Associates microBondapak C18) using a mobile phase composed of methanol and phosphate buffer at pH 6.3. Vancomycin was quantitatively recovered from human plasma samples (>96%) with high values of precision. The separation was completed within 27 min. The calibration curve was linear over the range from 5 to 1,000 ng/mL with the detection and quantification limits of 2 ng/mL and 5 ng/mL, respectively. This method is suitable for the routine assay of plasma samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.