Abstract
For this direct colorimetry of urinary oxalate, commercially available oxalate oxidase (EC 1.2.3.4) is used. The urine is first diluted, to diminish the effect of interfering substances. Analytical recovery of oxalate from urines with five different oxalate concentrations (0.4 to 2.0 mmol/L) ranged from 92 to 109% (mean 99%). The within-day and between-day precision (CV) of the method for a wide range of oxalate concentrations averaged better than 10%. There is good correlation (r = 0.977) between this enzymatic method (y) and the chemical method of Hodgkinson and Williams (x) [Clin Chim Acta 36: 127-132, 1972], the regression equation being y = 1.014x + 0.061. Urines with added ascorbate give falsely increased results. The proposed method is inexpensive and simple to perform.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.