Abstract

Energy dispersive X-Ray fluorescence (EDXRFS) method is developed and standardized for the determination of uranium on routine basis in various process stream solutions, covering a vide range of concentrations from 0.1 to 400 g L−1, from an Uranium Extraction Plant at Nuclear Fuel Complex. The method has been applied to aqueous stream samples. Except for dilution, no much sample preparation was involved in the analysis and accordingly the experimental parameters were optimized. The calibration curve in the range of 0.1–10 g L−1 of U was drawn manually using synthetic standard solutions prepared from U3O8 powder and Lα (13.61 keV) line of uranium was used for the measurements. The results from EDXRFS method are compared with other methods and are found to be in good agreement. The EDXRFS measurements carried over a range of 0.1–350 g L−1 of uranium have shown a RSD of ±1–5%. Also, the limitations of reported methods in literature and the advantages of present method are highlighted in the paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.