Abstract

AbstractAnlysis of uranium (VI) by X‐ray fluorescence methods, using selective solvent extraction with tri‐n‐ octulphosphin‐oxide (TOPO) has been studied. The analysis of uranium in copper leach solutions has proved successful and results indicate that this method can be used for uranium levels doen to 3ppm. The coefficient of variation (CV) at a value of 30 ppm was calculated to be CV = 2.6%.