Abstract

A simple and fast method for the simultaneous determination of two mouldy compounds, 2,4,6-trichloroanisole (TCA) and 2,4,6-tribromoanisole (TBA), in cork by gas chromatography-mass spectrometry (GC-MS) was established. The analytes were extracted by ultrasonic extraction with methanol, and purified then by solid phase extraction using primary secondary amine (PSA) as solid phase. After concentrating, the sample was analyzed by GC-MS and quantified by the external standard method. The linear ranges were from 10 microg/L to 10 000 microg/L for TCA and TBA, the correlation coefficients (r2) of the calibration curves were above 0.99. The recoveries and the relative standard deviations (RSDs) of TCA and TBA in different kinds of corks were investigated. The recoveries ranged from 88.4% to 97.6% with the RSDs between 1.02% and 4.58% (n = 6). The limits of detection (LODs) were 12 microg/L for TCA and 18 microg/L for TBA, and the limits of quantification (LOQs) were 40 microg/L for TCA and 50 microg/L for TBA. The method is suitable to the determination of TCA and TBA in corks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.