The determination of tungsten in steels and alloys with the three-electrode direct current plasma (DCP) “spectrajet” was investigated. The relative intensities of 17 spectral lines of tungsten and several possible interfering lines of its concomitants are listed. WI 400.875 nm is normally best suited, because the background produced by iron and acid is low and easily compensated for by using a blank solution. The limit of detection c l in steels is 2 × 10 −3% W for 10 mg ml of sample. The line W I 407.436 nm is preferred in the presence of much titanium. For the purpose considered the DCP provides about the same power of detection as the more expensive inductively coupled plasma (ICP), but its observation zone is less well buffered against influences from the matrix.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.