Abstract

Objective: To establish the method for determination of tungsten and insoluble compounds in the air of workplace by inductively coupled plasma optical emission spectrometry (ICP-OES) . Methods: The tungsten and insoluble compounds were digested by microwave digestion apparatus using nitric acid and hydrofluoric acid, detected by ICP-OES. Results: The linearity of tungsten and insoluble compounds (as tungsten) were good at the range of 0.16-100.0 μg/ml, the minimum quantitation concentration was 0.11 mg/m(3), the recovery was ranged from 81.9%-97.1%, the RSD of intra-and inter-batch precision were 0.5%-2.5% and 1.2%-4.4%, respectively. Conclusion: The determination method meet the requirement of analysis and apply to the determination of tungsten and insoluble compounds in the air of workplace.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call