Abstract

This work describes a new analytical procedure for trace vanadium by graphite furnace atomic absorption spectroscopy coupled to cloud point extraction (CPE) as the separation-preconcentration method. The CPE behavior of vanadium using methylene blue as complex agent and Triton X-100 as a surfactant was investigated systematically. Under the optimized conditions, the detection limit was 0.7 ng · mL−1, and the relative standard deviation was 4.3% for vanadium (c = 50.0 ng · mL−1, n = 5). The recovery of vanadium was in the range of 98.9–102.8%. The method was applied to the analysis of vanadium in certified reference materials and real samples. The results obtained were in good agreement with the certified values.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call