Abstract
Selenium is determined in the ng g −1 to μg g −1 range in biological and environmental samples. A wet digestion procedure was optimized with respect to volatility losses and recovery yields, by using 75Se metabolically incorporated into rat organs. Selenium is preconcentrated from the digestion liquid by a two-step reduction with 4 M HCl and ascorbic acid. The colloidal selenium formed is adsorbed on activated carbon and filtered on a Nucleopore membrane for measurement by energy-dispersive x-ray fluorescence. Almost complete recovery was obtained, and the detection limit was 20 ng, corresponding to 10 ng g −1 for a 2-g sample. Biological reference materials were analyzed with satisfactory results, and the accuracy of the method was good.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.