Abstract
A procedure of Se(IV) determination by anodic stripping voltammetry using two deposition and stripping steps at gold electrodes was proposed. A well-defined stripping peak of selenium was obtained at potential 0.9 V. The optimization of parameters influencing the selenium peak current including both deposition and stripping steps was performed. A linear relationship was observed between the Se(IV) peak current and its concentration in the range from 5×10−9 to 1×10−7mol L−1. The limit of detection was found to be 8.5×10−10mol L−1. Repeatability of the method determined as RSD % for Se(IV) concentration of 5×10−8mol L−1 was 4.3% (n=7). The proposed procedure was used for Se(IV) determination in certified reference material and natural water samples and acceptable results and recoveries were obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.