Abstract

In the present work a method for the determination of perchlorate trace levels by total reflection X-ray fluorescence (TXRF) is introduced. Perchlorate anions were concentrated on anion-selective membranes that had been prepared on the surface of TXRF quartz reflectors. Various complexation substances were used in the membranes. The reflectors were immersed in water solutions containing nanogram per milliliter (ppb) concentrations of perchlorate. After this step, the reflectors were taken out of the solution and they were analyzed by TXRF, using a copper X-ray tube and helium flow on the target (to lower the argon peak which is present in the air). The effects of various experimental parameters were examined, and the possibility of discrimination between chloride and perchlorate anions was shown. Minimum detection limits lower than 1 ng/mL and good linearity at the concentration range of 1-50 ng/mL were achieved. The method is applicable for the analysis of perchlorate in drinking water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.