Abstract

A method for the determination of trace organochlorine pesticides (OCPs) in soil using isotope dilution and high resolution gas chromatography-high resolution mass spectrometry (ID-HRGC-HRMS) was developed. The sample was extracted by accelerated solvent extractor (ASE) and cleaned-up by a Florisil solid phase extraction (SPE) cartridge. The analytes were separated by HRGC on a DB-5MS column (30 mx 0.25 mm x 0.25 microm) and determined by HRMS. The identifications of OCPs were based on the retention time of 13C-labelled standard and the abundance ratio of the two exact mass-to-charge ratios. The quantitative analysis was performed using the ratios of the integrated areas of the 13C-labelled standards. This method has the recoveries ranging from 77.3% to 114.5% and the relative standard deviations (RSD) less than 10.81% (n=5). The limits of detection (LODs) of this method for all OCPs were lower than 0.04 pg/g. The results indicated that the method is rapid, selective and sensitive for precise determination requirements of organochlorine pesticides at trace level in soil.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.