Abstract
The determination of trace lead in a CaCO 3 matrix by isotope dilution coupled anion-exchange and quadrupole-based inductively coupled plasma-mass spectrometry was established. The matrix of CaCO 3 seriously affected the determination by clogging the cones of the instrument was removed by the anion-exchange resin column. Effects such as dead time, mass bias, dwell time and the amount of isotopic agent spiked were discussed and corrected. The detection limit estimated by 3× standard deviation of the procedure blank was 2.94 ng g −1. The results indicated that the lead concentration in all samples tested was far less than 3%, the legislated limit. The recovery mostly ranged from 90 to 110%. The relative standard deviation was found to be <1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.