Abstract
A new method for the determination of trace lead was developed by dispersive liquid–liquid microextraction preconcentration and graphite furnace atomic absorption spectrometry. In the proposed approach, 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) was used as a chelating agent, and carbon tetrachloride and ethanol were selected as extraction and dispersive solvents. Some factors influencing the extraction efficiency of lead and its subsequent determination, including extraction and dispersive solvent type and volume, pH of sample solution, concentration of the chelating agent, and extraction time, were studied and optimized. Under the optimum conditions, the enrichment factor of this method for lead was reached at 78. The detection limit for lead was 39 ng L −1 (3σ), and the relative standard deviation (RSD) was 3.2% ( n = 7, c = 10 ng mL −1). The method was successfully applied to the determination of trace amounts of lead in human urine and water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.