Abstract

A method is described for the direct determination of iron in high purity sodium fluoride using graphite furnace atomic absorption spectrometry. Interferences caused by the matrix are investigated. It is shown that the ashing temperature can be increased to 1 400°C and matrix interferences eliminated, the sensitivity of iron increased in 1. 27 fold by the addition of nickel nitrate. The method is applied to the determination of iron in sodium fluoride and satisfactory results are obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.