Abstract
A sample dissolution-concentration technique for the determination of trace elements in fused silica by spark-source mass spectrometry (s.s.m.s.) is described. The fused silica is dissolved by ultra-pure hydrofluoric acid, and most of the silicon tetrafluoride formed is then removed at 80–100° C, leaving behind the impurities in the solution as fluorides. After addition of ultrahigh-purity graphite and internal standards (Tl and Rb), the mixture is dried to give a sample suitable for pressing into electrodes. The method not only concentrates the trace impurities but also reduces the interferences originating from silicon oxide and carbide species in the mass spectra and thus allows the determination of a larger number of elements at p.p.b. levels. More than 30 trace impurities in 6 fused silica boules and J.T. Baker's Ultrex silicon dioxide are determined and compared with data obtained by other methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.