Abstract

A sequential injection-square-wave anodic stripping voltammetry (SIA-SWASV) is proposed for the simultaneous determination of Pb(II), Cd(II) and Zn(II), employing an in situ plated bismuth film screen-printed carbon electrode (Bi-SPCE) as a working electrode and hydrochloric acid as a supporting electrolyte. Bi(III) and analyte metal ions were on-line deposited onto a SPCE at -1.4 V vs. Ag/AgCl for 180 s. At a stopped flow, a square-wave voltammogram was recorded from -1.3 to 0 V vs. Ag/AgCl. The experimental conditions were optimized. Under the optimum conditions, the linear ranges were 0 - 70 microg L(-1) for Pb(II) and Cd(II), and 75 - 200 microg L(-1) for Zn(II). The limits of detection (S/N = 3) were obtained at concentrations as low as 0.89 microg L(-1) for Pb(II) and 0.69 microg L(-1) for Cd(II) for a 180-s deposition time. The proposed method was applied to the determination of Pb(II), Cd(II) and Zn(II) in water samples with satisfactory results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call