Abstract
An X-ray fluorescence method (XRF) is presented that allowed low detection limits (at the 0.1-23 ng mL(-1) level) to be obtained for Cr, Mn, Fe, Ni, Zn, Sr, Pb, Bi and Br in water. The samples were prepared using a thin layer method. Trace elements were determined via the calibration curve and standard addition. Absorption effects and inhomogenities in prepared samples were checked for using the emission-transmission method and internal standards, respectively. The results from the XRF method were compared with the results from the inductively coupled plasma atomic emission spectrometry method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.