Abstract

Traces of the specified elements are separated from 1 g of manganese (II), using a 30- g column of AG50W-X8 cation-exchange resin and mixtures of hydrochloric acid and acetone as eluents. The trace elements are separated into three groups and are determined by atomic absorption spectrometry, except uranium for which spectrophotometry is used. Recoveries for 10 μg amounts (20 μg for gallium) vary between 94% (for gallium) and 103% (for uranium). A combined elution curve, results for the analysis of synthetic mixtures and for the determination of ten trace elements in samples of manganese metal, chloride and dioxide are presented.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call