Abstract

Epithermal neutron activation analysis (ENAA) was applied to measure quantitatively Cd and other elements in bone samples from control and Cd-fed rats. This method was found to be non-destructive to the bone samples, with no sign of “radiolytic charring” and was sensitive enough to detect and quantify Cd in bone samples at normal levels for mammals (viz. 0.5–1.0 μg/g) and higher. Two different thermal neutron shield materials were utilized, namely cadmium and boron. The boron shield resulted in a 27% improvement in the detection limit of Cd in bone. The accuracy of ENAA for Cd was assessed by intercomparison with electrothermal atomic absorption spectrophotometry (ETAAS), and the results were in fair agreement (±23%) with those from ENAA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.