Abstract

A sensitive and selective procedure is presented for the voltammetric determination of lead. The procedure involves an adsorptive accumulation of lead L-3-(3,4-dihydroxyphenyl)alanine (LDOPA) on a hanging mercury drop electrode, followed by a stripping voltammetric measurement of reduction current of an adsorbed complex at –0.15 V (vs Ag|AgCl). Optimum conditions for lead analysis include pH 8.5, 80 μM LDOPA and accumulation potential –0.15 V (vs Ag|AgCl). The peak currents are proportional to the lead concentration 1–300 nmol l–1 with a detection limit of 0.6 nmol l–1 and accumulation time 60 s. The method was used for the determination of lead in blood, dry tea and also in waters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.