Abstract

A new online flow injection spectrophotometric method for the determination of trivalent chromium was developed. This method is based on the property of trivalent chromium to be a catalyst for the oxidation of Indigo Carmine (IC) with potassium periodate and to lose its color in the presence of ethylenediaminetetraacetic acid and sodium tripolyphosphate. It was shown that Tween-20 serves as an additional accelerator of the catalytic oxidation. The linear dynamic range of the determination of Cr(III) was 1–40.0 μg/L, while the limit of detection was 0.05 μg/L. The correlation coefficient r was 0.998, while the relative standard deviation for 5 μg/L of the Cr(III) solution was 3.83%. The feasibility of this method was checked by its application to trivalent chromium determination in real water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.