Abstract
A novel method based on the coupling of membrane-supported headspace single-drop microextraction with gas chromatography‒mass spectrometry (GC–MS) is developed for the determination of chlorobenzenes in water samples. For the determination of five chlorobenzenes, a 15 μL toluene microdrop was placed inside the plastic membrane and exposed for 10 min for headspace extraction while stirring at 1000 rpm. The microdrop was then picked up by a microsyringe and directly injected into the injector block of the GC–MS instrument. Under the optimized operation conditions, the calculated calibration curves gave a high level of linearity for all targets with correlation coefficients range from 0.9945 to 0.9987. The limits of detection range from 0.01 to 0.05 μg/L and the RSDs for most of chlorobenzenes were below 7%. The method is simple, sensitive, and stable for single drop microextraction. Its applicability is demonstrated by the determination of chlorobenzenes in tap water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.