Abstract

A method is developed for the determination of trace mercury in biological samples using photo chemical vapor generation (PVG) and isotope dilution inductively coupled plasma mass spectrometry (ID ICPMS) detection. Biological tissues were solubilized in formic acid. Subsequently, the sample solutions were exposed to an ultraviolet (UV) source for the reduction of mercury into vapor species prior to ICPMS measurements. The formic acid served not only as a tissue solubilizer in the sample preparation procedure, but also as a photochemical reductant for mercury in the PVG process. The problem arising from the opaque formic acid digested solution was efficiently solved by using ID method. The optimum conditions for sample treatment and PVG were investigated. A limit of detection (LOD) of 0.5 pg g(-1), based on an external calibration, provided 350-fold improvement over that obtained by utilizing conventional pneumatic nebulization sample introduction. Method validation was demonstrated by the determination of total mercury in several biological tissue certified reference materials (CRMs). The results were in good agreement with the certified values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.