Abstract

The most widely used method for newborn screening for homocystinuria (HCU) is a semi-quantitative bacterial inhibition assay for measuring methionine concentration in dried blood spots (DBS). Because this method has resulted in a number of missed cases due to many factors, we developed a high performance liquid chromatography (HPLC) method with fluorescence detection to measure total homocysteine (tHcy) in DBS which might be useful for newborn screening for HCU. One disk of DBS 3 mm in diameter was sonicated in 10 min. The extract was reduced with dithioerythritol and was derivatized with 4-aminosulfonyl-7fluoro-2,1,3-benzoxadiazole before injection into HPLC. This method showed good linearity (r = 0.996), precision (coefficient of variation range 2.7-5%), and excellent correlation coefficient between DBS and serum tHcy, both in control (r = 0.932) and patient samples (r = 0.952). By this method, the mean tHcy concentration in DBS of preterm newborns, full-term newborns, and adults was 1.4 +/- 1.0, 2.5 +/- 1.6, and 4.9 +/- 1.5 micro mol/L, respectively. The mean tHcy DBS concentration in two cases of cystathionine-beta-synthase deficiency and one case of 5,10-methylentetrahydrofolate reductase deficiency was 22.7 +/- 2.88, 29.3 +/- 1.90, and 41.3 micro mol/L, respectively. The present method, which is rapid, user friendly and reliable, seems applicable to newborn screening of HCU in place of methionine measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.