Abstract

Most newborn screening laboratories use CE-marked or FDA-approved test-kits, like in routine clinical chemistry. National regulations require only minimal evaluation from the customer, if the test-kits are used as specified by the manufacturer. The microtiter-based kit-concept is often based on the perception, that the laboratory always processes whole microtiter plates. However, in the daily routine, this is rather a rare exception, which leads to much higher costs per newborn, compared to the costs per assay in the test-kits. In addition the amount of wasted resources is quite high. Performance of the Neonatal Total Galactose kit from Perkin Elmer was tested. We have determined specificity, limit of detection (LOD), limit of quantitation (LOQ), intra and inter assay variation, recovery, stability of measuring signal and reagents. Results were also compared with the Astoria Pacific Spot Check System. In addition, we had (by chance) the opportunity to test 2 kits, which were already expired for more than 3 years. LOD was 165 - 306 μmol/L and LOQ 475 - 703 μmol/L, depending on the definition of LOD/LOQ. Mean recovery was 112.8%, intra assay CVs were 11.3, 7.3, 4.0, and 3.0, and inter assay CVs 28.7, 15.9, 7.8, and 9.3, at 220, 590, 1200, and 2060 μmol/L respectively. Reconstituted and mixed reagents must be used within some hours, and were unstable even if stored at -20℃. However, if the reconstituted galactose substrate reagent and galactose oxidase reagent were only mixed according to the daily requirements, and the rest stored separately at -20℃, they were stable for at least 12 days. The performance of the expired test-kits did not differ from the others. The performance of the Total Galactose kit is comparable to other tests used for newborn screening. However, we could significantly reduce the costs per newborn and reduce unnecessary production of waste, by thorough validation and modification of the assay procedures.

Highlights

  • Newborn Screening for Galactosemia was introduced in the 1970s [1]

  • National regulations require only minimal evaluation from the customer if the test-kits are used as specified by the manufacturer

  • We have thoroughly tested the performance of the Neonatal Total Galactose kit from Perkin Elmer

Read more

Summary

Introduction

Newborn Screening for Galactosemia was introduced in the 1970s [1]. While the classical “Beutler-Test” can only detect newborns with classical galactosaemie [2], the determination of total galactose [3,4] will detect newborns with galactokinase deficiency and uridine diphosphate galactose-4-epimerase deficiency.Most newborn screening laboratories use CE-marked or FDA-approved test-kits, like in routine clinical chemistry. In the daily routine, this is rather a rare exception, which leads to much higher costs per newborn, compared to the costs per assay in the test-kits. Performance of the Neonatal Total Galactose kit from Perkin Elmer was tested. Limit of detection (LOD), limit of quantitation (LOQ), intra and inter assay variation, recovery, stability of measuring signal and reagents.

Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.