Abstract

AbstractA new simple, rapid and sensitive spectrophotometric flow injection analysis (FIA) method is developed for the determination of chromium(VI) based on its reaction with 3,5‐Br2‐PADAP and SCN°‐ in a strongly acidic medium. This produces an unstable violet product with a maximum absorption at 605 nm. As the product formed is unstable, FIA is an ideal method to reproducibly monitor the transient signal. The calibration curve is linear over 1.0 × 10–6 to 7.0 × 10–5 mol l–1. The correlation coefficient of the calibration curve is 0.9998, the detection limit is 5.0 × 10–7Cr2O72‐ mol l–1 and sample throughput is 80 h–1. The method has been successfully applied to the determination of chromium in steel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.