Abstract

AbstractWe described here a new method for the determination of total calcium in plasma. The method is based on the precipitation of calcium with excess oxalate and the measurement of residual oxalate by flow injection analysis with Ru(bpy)32+ electrochemiluminescent detection. It has the advantages of extremely stable reagent, user‐friendly instrument, high selectivity, good analytical recovery, wide dynamic range, and nice correlation with atomic absorption spectroscopy. The calibration plot for calcium is linear over a concentration range from 0.5 mmol L−1 to 4.8 mmol L−1, which is wider than those obtained by most other methods. The analytical recoveries for plasma calcium are 98.4–101.2% with coefficients of variation (CVs) of 1.96–2.52%. The within‐day CVs range from 0.76% to 0.95%, and between‐day CVs were from 1.12% to 1.46%. The time for each injection is one minute. Because the proposed method can be readily carried out on increasingly popular instruments for Ru(bpy)32+ ECL immunoassays and DNA probe assays, Ru(bpy)32+ ECL method is suitable for routine clinical analysis of calcium.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.