Abstract
A square-wave polarographic method for the determination of the amino acid thyroxine in urine, after isolation from the matrix on a C18 Sep-Pak cartridge and subsequent elution through a column packed with Amberlite IRA-400 anionic resin, was developed. The optimal working conditions for the application of the method were found to be pH 10, provided by a carbonate buffer, and a potential of -0.690 V during 10 min. The equilibration time was applied during 5 s, and potential scans were performed at a square-wave amplitude of 100 mV and a scan rate of 300 mV s-1, with a square-wave frequency of 100 Hz. The measuring-system response was linear over the thyroxine concentration range from 10 to 60 ng ml-1 and the detection limit achieved was 2.71 ng ml-1. The relative error and relative standard deviation obtained were 1.18 and 4.67%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: The Analyst
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.