Abstract

The level of three alkylphenols (APs) 4-nonylphenol, 4-octylphenol and 4-tert-octylphenol was monitored in various water samples using a new developed and validated HPLC method. The HPLC method employed a column with a mixed-mode stationary phase (C18/SCX) using a mobile phase of water to methanol = 15:85 (v/v) delivered at a flow rate of 1 mL/min at room temperature. Both diode array, DAD and fluorescence, FLD detectors were employed. The method is linear when APs concentration ranged from 0.025 to 0.5 μg/mL with a DAD detection at 279 nm and between 0.0008 and 0.1 μg/mL when the detection was made by FLD (excitation at 220 nm, emission at 315 nm). The limit of detection and limit of quantitation for alkylphenols were found to be 5 and 15 ng/mL, respectively (detection by DAD). The method was employed with good results for the determination of APs in the presence of polycyclic aromatic hydrocarbons in various water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.