Abstract

This work evaluates the wavelength reproducibility and accuracy of a sequential scanning plasma emission system. The long-term thermal and thermomechanical errors are found to be in the range of a few picometers. The short-term instability and mechanical instability are found to contribute very little error to wavelength positioning. The wavelength calibration algorithm is described. Its use improves the accuracy by a factor of 5 to 10 by comparison with results from the use of the diffraction grating equation alone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.