Abstract

There are many dynamic methods for measuring the volumetric mass transfer coefficient. The "gas out-gas in" method can directly determine the volumetric mass transfer coefficient in a bioreactor system and provide estimates of the volumetric microbial oxygen uptake rate and the average oxygen saturation concentration at the gas-liquid interface. The errors on these parameters are large if the dissolved oxygen probe response time is not considered. For reliable measurements, deconvolution of the oxygen probe measurements must be made. (c) 1995 John Wiley & Sons, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.