Abstract
We describe a methodology for determining the volume scattering function β(ψ) of aqueous particle suspensions from measurements with a laboratory multi-angle light scattering instrument called DAWN (Wyatt Technology Corporation). In addition to absolute and angular calibration, the key component of the method is the algorithm correcting for reflection errors that reduce the percent error in β(ψ) from as much as ~300% to <13% at backward scattering angles. The method is optimized and tested with simulations of three-dimensional radiative transfer of exact measurement geometry including the key components of the instrument and also validated experimentally using aqueous suspensions of polystyrene beads. Example applications of the method to samples of oceanic waters and comparisons of these measurements with results obtained with other light scattering instruments are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.